The Fluke 726 PrecisionMultifunction Process Calibratoris designed specifically for the Process industry with broad workload coverage, calibration power and unsurpassed accuracy in mind. The 726 measures and sources almost all process parameters and can calibrate almost anything. The 726 will also interpret results without the help of a calculator and store measurement data for later analysis.
Includes:
One set of AC72 alligator clips
One set of stackable test leads
One set of TL75 testleads
User manuals
NIST- traceable calibration report and data
The Fluke 726 Precision Multifunction Process Calibrator is precise and powerful, yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.
More precise measurement and calibration source performance, accuracies of 0.01%.
Two separate channels; measure, source and view process signals simultaneously
Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters
Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters
Measure or source pressure* using any of 29 Fluke 700Pxx Pressure Modules
Source mA with simultaneous pressure measurement to conduct valve and I/P tests
Integrated pressure switch test allows you to capture the set, reset and deadband of a switch
Memory storage for up to 8 calibration results, return stored calibration data from the field for later analysis
Transmitter error % calculation, interpret calibration results without a calculator
Perform fast linearity tests with auto step and auto ramp features
Power transmitters during test using 24 V loop supply and simultaneous mA measurement
Store frequently-used test setups for later use
Frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
HART mode inserts 250 ohm resistor in mA measure and source for compatibility with HART instrumentation
Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement.
New voltage input protection design for improved reliability
Three-year warranty
Size
200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in)
Weight
650 g (23 oz)
Batteries
4 AA alkaline batteries
Accuracy
Measurement Accuracy
Voltage DC
30.000 V
0.01% + 2 mV (upper display)
20.000 V
0.01% + 2 mV (lower display)
90.00 mV
0.01% + 10 μV
-10.00 mV to 75.00 mV
0.025% + 1 count (via TC connector)
Current DC
24.000 mA
0.01%+ 2 mA
Resistance
0.0 to 4000.0 Ω
± 0.015%
Frequency
2.0 to 1000.0 CPM
0.05% + 0.1 cpm
1.0 to 1000.0 Hz
0.05% + 0.1 Hz
1.00 to 15.00 kHz
0.05% + 0.01 kHz
Sensitivity
1 V peak-to-peak-minimum
Pressure
Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC
100.00 mV
0.01% + 10 μV
20.000 V
0.01% + 2 mV
-10.00 mV to 75.00 mV
0.025% + 1 count (via TC connector)
Current DC
24.000 mA (Source)
0.01% + 2 μA
24.000 mA (Simulate)
0.01% + 2 μA
Resistance
Range, accuracy (percent of reading + floor) and excitation current
5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA
Frequency
2.0 to 1000.0 CPM
0.05%
1.0 to 1000.0 Hz
0.05%
1.0 to 10.0 kHz
0.25%
10.0 to 15.0 kHz
0.5%
Waveform
5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure and source accuracy
Cu10
1.8°C
NI-120
0.15°C
PT-100 (385)
0.45°C
PT-100 (392)
0.35°C
PT-100 (391-JIS)
0.35°C
PT-200 (385)
0.95°C
PT-500 (385)
0.55°C
PT-1000 (385)
0.35°C
RTD Resolution
0.01°C (0.01°F)
Cu10.01 C
0.1°C (0.1°F)
J
0.4°C
K
0.5°C
T
0.4°C
E
0.4°C
R
1.4°C
S
1.4°C
B
1.4°C
L
0.45°C
U
0.45°C
N
0.6°C
XK
0.4°C
BP
1.1°C
TC Resolution
J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F